Options
Chhabra, Meenu
Loading...
Preferred name
Chhabra, Meenu
Alternative Name
Chhabra, M.
Main Affiliation
Scopus Author ID
24328801100
Researcher ID
CKB-3624-2022
Now showing 1 - 1 of 1
- PublicationGraphene oxide modified screen printed electrodes based sensor for rapid detection of E. coli in potable water(2024)
;Vandana Kumari Chalka ;Khushi Maheshwari; ; A sensitive and rapid electrochemical sensor has been developed to detect Escherichia coli (E. coli) bacteria that is crucial for ensuring safe drinking water. E. coli significantly contributes to waterborne contamination, driven by overexploitation and insufficient cleanliness around water bodies. This work incorporates synthesis and characterization of graphene oxide (GO), sensor preparation, and sensor testing in the presence of varying E. coli dilutions via H2O2 decomposition. GO is used as a sensing layer and drop casted on the working electrode of screen printed electrode (SPE). The sensor is exposed to different bacterial solutions using varied bacterial concentrations and fixed percent solution of H2O2. The interface of GO and bacterial solution results in a change in potential. The cross-sensitivity tests have also been performed in the presence of chemical compounds found in wastewater samples, Pseudomonas aeruginosa and Citrobacter youngae. The sensor demonstrates a detection limit of 2.8 CFU/mL, with a sensitivity of 5.1 mV-mL/CFU, fast response time and excellent repeatability when tested with E. coli. Its performance has also been assessed by comparing the sensing results for regular tap water, reverse osmosis (RO) water, and deionized water samples. This work paves the way for developing a chip-based system to detect E. coli in water.